27 November 2018: Original Paper
Evaluation Using an Isolated Reperfusion Model for Porcine Liver Donated After Cardiac Death Preserved with Oxygenated Hypothermic Machine Perfusion
Ryo Yoshikawa ABCDEF 1, Naoto Matsuno ABDEFG 1,2,3*, Noriyuki Morito ABCDF 1, Mikako Gouchi BCD 2, Masahide Otani D 2, Hiroyuki Takahashi D 2, Tatsuya Shonaka D 2, Yuji Nishikawa BCD 4, Shin Enosawa AD 3, Toshihiko Hirano AD 5, Hiroyuki Furukawa AD 2, Hiromichi Obara ABDEFG 1,3
DOI: 10.12659/AOT.910008
Ann Transplant 2018; 23:822-827
Abstract
BACKGROUND: Machine perfusion techniques offer a solution to the serious organ shortage. However, to assess the effects of machine perfusion, many detailed studies are required. In this study, an ex vivo reperfusion model using diluted autologous blood was confirmed to evaluate the utility of machine preservation for livers donated after cardiac death (DCD). In particular, beneficial effects of the oxygenated hypothermic machine perfusion (HMP) for DCD porcine livers are evaluated.
MATERIAL AND METHODS: Porcine livers were procured under warm ischemia time (WIT) of 60 min. The livers were preserved by hypothermic machine perfusion (HMP) or static cold storage (CS) for 4 h. After the preservation, the livers were perfused for 2 h using the ex vivo reperfusion model with diluted blood oxygenated by a membrane oxygenator at 35–38°C.
RESULTS: At 2 h of ex vivo reperfusion with 60 min of warm ischemic time (WIT), the portal vein pressure for CS was higher than HMP (18.8±15.9 vs. 7.5±3.9 [mmHg] in 60 min). Furthermore, LDH in CS was higher than HMP (528.5±149.8 vs. 194.1±32.2 [IU/L/100 g liver] in 60 min. P<0.05). Lactate after CS (60) was significantly higher than HMP (60) (8.67±0.39 vs. 5.68±0.60 [mmol/L] at 60 min. p<0.01).
CONCLUSIONS: The ex vivo reperfusion model can be used to evaluate the utility of machine perfusion. Advantages of HMP for DCD livers are evaluated with this model.
Keywords: Liver Transplantation, Organ Preservation
Background
The current criterion standard method of organ preservation, simple cold storage (CS), has been widely accepted. Liver machine perfusion preservation (MP) techniques for
Hypothermic machine perfusion (HMP) provides continuous circulation of preservation solution and metabolic substances. A washout effect removes and dilutes waste products from direct endothelial/parenchymal contact due to stabilization of the microcirculation. There is strong evidence that renal HMP improves the early graft function [1] and improves the tissue ATP concentrations upon reperfusion [2]. HMP also allows for pharmacological manipulation and the pretransplant assessment of grafts [2] and has enhanced the use of ECD kidneys [1]. Furthermore, recent reports suggest a long-term graft survival benefit for HMP kidneys [2,3].
A clinical pilot study was performed with normal livers compared with CS alone and with CS followed by HMP [4], showing that the rates of early allograft dysfunction, postoperative transaminases, bilirubin, and hospital stay were significantly lower in the CS+HMP group compared with the CS-only group. The same authors later performed HMP in ECD livers and reported the safety and efficacy of HMP in such a scenario [5]. In addition, the utility of normothermic machine perfusion (NMP) recently showed the possibility of MP, showing that the rate of graft injury, as measured by hepatocellular enzyme release and organ discard was lower and mean preservation time was longer in the NMP group compared with the CS group [6].
However, the optimum temperatures and conditions for organ preservation remain controversial. Studies on MP have been performed in the following 3 temperature categories: hypothermic (6–10°C), subnormothermic (20–23°C), and normothermic (37°C) [6–8]. However, more detailed large-animal studies with a sufficient number of subjects are required to confirm the utility of these techniques and evaluate the preservation results. Furthermore, from the perspective of animal ethics, the development and implementation of alternatives to animal use should be considered when designing experiments.
In the present study, an isolated
Material and Methods
ANIMALS:
Experiments were performed on porcine liver grafts (domestic, female, cross-bred Large-Yorkshire, Landrace, and Duroc pigs, 2 to 3 months old, 20 kg; Tokyo Experimental Animals Supply Co., Ltd., Tokyo, Japan). These harvesting procedures were conducted under inhalation anesthesia with isoflurane (Forane VR; Abbot Japan, Tokyo, Japan), and hemorrhagic shock and the removal of ventilation induced cardiac arrest. The Institutional Animal Ethics Committee of the Clinical Research Center, National Center for Child Health and Development (Permit no. 2010-010 A) approved all experimental procedures.
PERFUSION PRESERVATION MACHINE:
The grafts were perfused using an original MP system (Figure 1) [9]. This system consists of the hepatic artery (HA) and portal vein (PV) perfusion circuits. Each circuit consists of a roller-type pump (MasterFlex7520-40; Cole-Parmer, IL, USA), an electrical flow meter (VN05; Aichi Tokei, Japan for PV; FD-SS02; Keyence, Osaka, Japan for HA), a ceramic capacitance pressure sensor (KL76; Nagano Keiki, Nagano, Japan), and a custom-made air trap. An oxygenator (HP0-06 H-C; Senko Medical Instrument, Tokyo, Japan) was installed in the circuit for the PV and HA. Two optical oxygen sensors (NeoFox; Ocean Optics, FL, USA) were installed to measure the oxygen concentration of the perfusate in the PV circuit. The perfusate temperature was controlled by a heat exchanger and ice-cold water. The flow rate of the PV was 0.22 mL/min/g and HA was 0.06 mL/min/g. The oxygen consumption (mg/min/100 g liver) of the organ was calculated as normalized values with the mass of the organ (mg/min/100 g liver) from the difference in the oxygen concentration at the inlet port of the PV circuit and the outlet port positioned near the hepatic vein (HV).
OXYGENATION:
Oxygenated perfusate supplied oxygen to the liver grafts during MP under HMP conditions using an oxygenator. The oxygen concentrations were controlled with the blending volume conditions of oxygen and air using the air blender of the oxygenator. The condition of the oxygen concentration was maintained to hold the minimum dissolved oxygen concentration for the calculation of the oxygen consumption at the outlet port. The oxygen consumption (mg/min/100 g liver) of the organ was calculated as normalized values with the mass of the organ (mg/min/100 g liver) from the difference in the oxygen concentration at the inlet port of the PV circuit and the outlet port positioned near the HV.
EXPERIMENTAL DESIGN:
Grafts were procured under a warm ischemia time (WIT) of 0 or 60 min and were preserved under CS or HMP for 4 h. The perfusate used a modified gluconate University of Wisconsin solution. We used 4 perfusion conditions in the present study (Figure 2):
The liver grafts were treated for 2 h using an isolated liver perfusion model, with the outcomes compared among 4 groups.
ISOLATED REPERFUSION MODEL (IRM):
Following the preservation, all organs were rinsed with 500 mL of cold Euro-Collins solution for 20 min, and subsequently exposed at room temperature for 10 min without perfusion to imitate the slow rewarming of the graft during surgical implantation
The livers were then reperfused with oxygenated diluted autologous blood at 38°C. The autologous blood was diluted with saline (40%), dextran (20%), calciol (5%), sodium bicarbonate, heparin sodium, and potassium. The hematocrit was maintained at around 10–12%. The oxygenator was regulated to achieve physiological blood gas values (pO2, approximately 150–200 mmHg; pCO2 approximately 30–50 mmHg). HA perfusion was set at 80 mmHg and automatically maintained by a roller pump connected to a pressure sensor placed in the inflow line immediately before the arterial cannula. Perfusion of the PV was maintained in a flow-constant manner (0.8 mL/min/g).
THE VIABILITY ASSESSMENT DURING MP:
Aspartate aminotransferase (AST) and lactate dehydrogenase (LDH) were taken from the perfusate every 30 min and analyzed to determine the viability of the preserved liver grafts using standard biochemical methods (clinical grade). Blood gas and lactate concentrations in perfusate were measured using the iStat program (Abbot, NJ, USA).
STATISTICAL ANALYSES:
Statistical analyses were performed using Microsoft Excel 2013 (Microsoft, WA, USA). The results are presented as mean ± standard deviation (SD). After proving the assumption of normality, differences between groups were tested by
Results
VASCULAR PRESSURE OF THE PV AND HA DURING IRM:
The PV pressure (PVP) and HA pressure (HAP) under each perfusion condition during IRM are shown in Figures 3 and 4, respectively. The HAP after CS (0) was higher than that after HMP (0) (151.0±17.9 vs. 50.4±41.9 mmHg at 30 min and 143.8±27.4 vs. 47.0±37.2 mmHg at 60 min after reperfusion). Furthermore, the HAP after CS (60) was significantly higher than that after HMP (60) (227.0±25.6 vs. 115.8±15.3 mmHg at 60 min and 171.5±27.2 vs. 109.8±9.8 mmHg at 120 min after reperfusion).
In addition, the PVP after CS (60) was higher than that after HMP (60) (18.8±15.9
AST, LDH, AND LACTATE DURING IRM:
The AST and LDH levels during IRM after CS (0), HMP (0), CS (60) and HMP (60) are shown in Figures 5 and 6. The AST level after CS (60) was significantly higher than that after HMP (60) at every point after reperfusion (# p<0.05, vs. IRM after CS (60) and HMP (60); Figure 6). Even at WIT 0 min, a significant difference was noted in the AST level between CS (0) and HMP (0) at every point after IRM (136.4±25.5 vs. 69.1±6.7 IU/L/100 g liver at 60 min and 218.9±27.4 vs. 120.2±41.4 IU/L/100 g liver at 120 min after reperfusion). Furthermore, the LDH level with IRM after CS (60) was significantly higher than that after HMP (60) (528.5±149.8 vs. 194.1±32.2 IU/L/100 g liver at 60 min and 576.8±150.6 vs. 226.4±50.5 IU/L/100 g liver at 120 min after reperfusion; p<0.05). The lactate level after CS (60) was significantly higher than that after HMP (60) at every point after reperfusion (8.67±0.39 vs. 5.68±0.60 mmol/L at 60 min and 11.48±0.84 vs. 7.29±0.67 mmol/L at 120 min, respectively; p<0.01; Figure 7).
Discussion
The utility of MP techniques for preserving the organ viability and recovering the organ function for organ transplantation has been reported previously [2,3], and the potential for HMP to increase the donor pool using marginal livers has also been reported [10–12].
Several different IRMs have been established to predict transplantation results [9,11,13]. In the present study, we established an IRM that reproduced ischemia-reperfusion injury to evaluate the utility of MP with DCD livers. With 2 h of reperfusion, the PVP, HAP, and LDH in the HMP (0) group remained relatively low, and better preservation conditions compared to CS (0) were noted. In contrast, in the CS (60) group, vascular pressure and LDH fluctuated drastically early after reperfusion, and obvious ischemia-reperfusion injury was noted. These differences in the perfusion flow dynamics and outflow of liver enzymes at the early stage of reperfusion might aid in rapid evaluation of organ function.
The continuous flow in MP allows for an influx of nutrients and oxygen while flushing cytokines, proteins, and toxins from the liver, thereby preventing the damage cascade buildup that occurs within a static preserved liver. The washout effects of MP and the continuous delivery of metabolic substances and antioxidants to the parenchyma and endothelium has been demonstrated to improve organ function [2,5].
Temperature modulation (e.g., preservation at normothermic and subnormothermic temperatures) offers a new means of improving organ preservation techniques with oxygenation to maintain graft viability and function. The advantages of normothermic MP (NMP) include physiological preservation [7]. A phase 1 clinical trial evaluating the efficacy of NMP [8] showed that NMP was safe as well as feasible, from retrieval to transplantation, including transportation. In addition, a clinical pilot study has recently reported the utility of normothermic machine perfusion (NMP) [6]. Several groups have examined the effects of setting the perfusion temperature to a subnormothermic temperature, which an intermediate temperature (20–33°C) between hypothermic and normothermic temperature conditions. Bruinsma et al. evaluated the efficacy of subnormothermic MP for
Shear stress occurs due to the reaction of oxygen with increased concentrations of intracellular redox-active iron, which triggers free-radical-mediated cell injury [17]. The present findings suggest that oxidative stress and endothelial shear stress can be decreased by regulating the perfusion temperature and oxygenation. Regarding oxygenation during HMP, the mechanism of tissue protection involves the reactivation of mitochondrial respiration, which occurs prior to reperfusion, when HMP oxidizes mitochondrial electron complexes. Mitochondrial electron overload on reperfusion can be prevented by the decreased production of reactive oxygen species [18]. Other studies have suggested that injury also occurs in the sinusoidal endoplasmic reticulum during both cold and warm ischemia [19,20]. However, extended periods of preservation without oxygenation may also be associated with certain disadvantages and limitations of endoplasmic stress and alterations of the sinusoidal endothelial cells [9,16]. Regarding aggressive oxygenation, our study showed that active oxygenation was not required for HMP.
In summary, HMP appears to protect against injury in 2 ways. First, oxygenation under hypothermic conditions protects against mitochondrial and nuclear injury by downregulating mitochondrial activity before reperfusion. Second, cold perfusion itself, under low pressure, prevents endothelial damage, independent of oxygen.
Conclusions
Our
References
1. Moers C, Smits JM, Maathuis MH, Machine perfusion or cold storage in deceased-donor kidney transplantation: N Engl J Med, 2009; 360; 7-19, pmid: 19118301
2. Kwiatkowski A, Wszola M, Kosieradzki M, Machine perfusion preservation improves renal allograft survival: Am J Transplant, 2007; 7; 1942-47, pmid: 17617857
3. Guarrera JV, Estevez J, Boykin J, Hypothermic machine perfusion of liver grafts for transplantation: technical development in human discard and miniature swine models: Transplant Proc, 2005; 37; 323-25, pmid: 15808631
4. Guarrera JV, Henry SD, Samstein B, Hypothermic machine preservation in human liver transplantation: The first clinical series: Am J Transplant, 2010; 10; 372-81, pmid: 19958323
5. Guarrera JV, Henry SD, Samstein B, Hypothermic machine preservation facilitates successful transplantation of “orphan” extended criteria donor livers: Am J Transplant, 2015; 15; 161-69, pmid: 25521639
6. Nasralla D, Coussios CC, Merqental H, A randomized trial of normothermic preservation in liver transplantation: Nature, 2018; 7703; 50-56
7. St Peter SD, Imber CJ, Lopez I, Extended preservation of non-heart-beating donor livers with normothermic machine perfusion: Br J Surg, 2002; 89(5); 609-16, pmid: 11972552
8. Bruinsma BG, Yeh H, Ozer S: Am J Transplant, 2014; 14(6); 1400-9, pmid: 24758155
9. Morito N, Obara H, Matsuno N, Regulated oxygenation of rewarming machine perfusion for porcine donation after cardiac death liver transplantation: Transpl Proc, 2016; 48; 1244-46
10. Obara H, Matsuno N, Shigeta T, Rewarming machine perfusion system for liver transplantation: J Med Devices, 2013; 7; 1-7
11. Compagnon P, Levesque E, Hentati H, An oxygenated and transportable machine perfusion system fully rescues liver grafts exposed to lethal ischemic damage in a pig model of DCD liver transplantation: Transplantation, 2017; 101(7); e205-13, pmid: 28403128
12. Raikumar R, Jassem W, Mergental HL, 2016; 16; 1779-87
13. Minor T, Efferz P, Fox M, Controlled oxygenated rewarming of cold stored liver grafts by thermally graduated machine perfusion prior to reperfusion: Am J Transplant, 2013; 13; 1450-60, pmid: 23617781
14. Furukori M, Matsuno N, Watanabe K, Subnormothermic machine perfusion preservation with rewarming for donation after cardiac death liver grafts in pigs: Transpl Proc, 2016; 48(4); 1239-43
15. Henry SD, Guarrera JV: Transplant Rev (Orlando), 2012; 26(2); 163-67, pmid: 22074785
16. Matsuno N, Uchida K, Furukawa H, Impact of machine perfusion preservation of liver grafts from donation after cardiac death: Transpl Proc, 2014; 46; 1099-110
17. Rauen U, Petrat F, Li T, De Groot H, Hypothermia injury/cold-induced apoptosis – evidence of an increase in chelatable iron causing oxidative injury in spite of low O2-/H2O2 formation: FASEB J, 2000; 14; 1953-64, pmid: 11023979
18. Du C, Fang M, Li Y, Smac, a mitochondrial protein that promotes cytochrome c-dependent caspase activation by eliminating IAP inhibition: Cell, 2000; 1; 33-42
19. Koetting M, Lüer B, Efferz P, Optimal time for hypothermic reconditioning of liver grafts by venous systemic oxygen persufflation in a large animal model: Transplantation, 2011; 91; 42-47, pmid: 21441852
20. Minor T, Manekeller S, Sioutis M, Dombrowski F, Endoplasmic and vascular surface activation during organ preservation: Refining upon the benefits of machine perfusion: Am J Transplant, 2006; 6; 1355-66, pmid: 16686759
In Press
Original article
Diagnostic Utility of FAR1 Methylation Levels in Hepatocellular Carcinoma Patients Undergoing Liver Transpl...Ann Transplant In Press; DOI: 10.12659/AOT.951568
Original article
Inferior Long-Term Outcome of Fatty Liver Allografts After Orthotopic Liver TransplantationAnn Transplant In Press; DOI: 10.12659/AOT.950589
Database Analysis
Identification and Validation of Liver Transplantation-Induced Acute Lung Injury Biomarkers Using a Bioinfo...Ann Transplant In Press; DOI: 10.12659/AOT.950289
Original article
Survival and Recurrence in Liver Transplant Patients With Intrahepatic Cholangiocarcinoma and Hepatocellula...Ann Transplant In Press; DOI: 10.12659/AOT.950997
Most Viewed Current Articles
24 Aug 2021 : Review article 18,372
Normothermic Machine Perfusion (NMP) of the Liver – Current Status and Future PerspectivesDOI :10.12659/AOT.931664
Ann Transplant 2021; 26:e931664
05 Apr 2022 : Original article 14,731
Impact of Statins on Hepatocellular Carcinoma Recurrence After Living-Donor Liver TransplantationDOI :10.12659/AOT.935604
Ann Transplant 2022; 27:e935604
22 Nov 2022 : Original article 14,244
Long-Term Effects of Everolimus-Facilitated Tacrolimus Reduction in Living-Donor Liver Transplant Recipient...DOI :10.12659/AOT.937988
Ann Transplant 2022; 27:e937988
29 Dec 2021 : Original article 13,752
Efficacy and Safety of Tacrolimus-Based Maintenance Regimens in De Novo Kidney Transplant Recipients: A Sys...DOI :10.12659/AOT.933588
Ann Transplant 2021; 26:e933588