Background: Right ventricular (RV) function is an important aspect of anesthesia management during orthotopic liver transplantation (OLT). Because of its geometrical complexity, assessment of RV dimensions with transesophageal echocardiography (TEE) is a difficult task. The aim of this prospective single-site study was to investigate the feasibility of intraoperative assessment of RV parameters based on reconstructive three-dimensional (3D) TEE and to compare the measurements to thermodilution-derived values acquired with a modified pulmonary artery catheter.
Material/Methods: Measurements were performed at four different time points during 30 OLT with 3D-TEE. At the same time comparative values of RV parameters were acquired with a fast-response thermistor pulmonary artery catheter.
Results: 3D reconstruction was feasible in all patients. RV dimensions measured with 3D-TEE averaged 119.4ml (±38.5 ml) for enddiastolic and 68.9 ml (±27.7 ml) for endsystolic volumes. The RV ejection fraction was 42.2% (±9.3%). The volumes obtained by thermodilution were 263.7 ml (±64.5 ml) enddiastolic and 159.3 ml (±47.5 ml) endsystolic, both significantly greater than by 3D-TEE, and the ejection fraction was found to be 39.5% (±8.4%). No correlation was found between the volumes or the function determined by either method.
Conclusions: Reconstructive 3D-TEE is a viable technique during OLT and leads to plausible RV parameters. However, no correlation was found with simultaneous measurements or parameters performed with thermodilution. However, based on our data cardiac output measurements by thermodilution appear reasonable. Due to both lack of agreement with 3D-TEE and extraordinary high RV volumes the question about the most valuable monitoring technique of RV dimensions and function during OLT can not finally be answered.
Keywords: Pulmonary artery catheter, transesophageal echocardiography, three-dimensional reconstruction, right ventricule, Orthotopic Liver Transplantation
560 4
In Press
17 May 2023 : Original article
Results of Liver Retransplantation After Rescue Hepatectomy: A Single-Center StudyAnn Transplant In Press; DOI: 10.12659/AOT.939557
10 May 2023 : Case report
Outcomes from a Single Transplant Center of 5 Pediatric Cases of Domino Liver Transplantation from Live Don...Ann Transplant In Press; DOI: 10.12659/AOT.939893
Most Viewed Current Articles
24 Aug 2021 : Review article
Normothermic Machine Perfusion (NMP) of the Liver – Current Status and Future PerspectivesDOI :10.12659/AOT.931664
Ann Transplant 2021; 26:e931664
26 Jan 2022 : Review article
Recurrence of Hepatocellular Carcinoma After Liver Transplantation: Risk Factors and Predictive ModelsDOI :10.12659/AOT.934924
Ann Transplant 2022; 27:e934924
29 Dec 2021 : Original article
Efficacy and Safety of Tacrolimus-Based Maintenance Regimens in De Novo Kidney Transplant Recipients: A Sys...DOI :10.12659/AOT.933588
Ann Transplant 2021; 26:e933588
15 Mar 2022 : Case report
Combined Liver, Pancreas-Duodenum, and Kidney Transplantation for Patients with Hepatitis B Cirrhosis, Urem...DOI :10.12659/AOT.935860
Ann Transplant 2022; 27:e935860